Description
The E201-C-9 pH Sensor Kit is an analog pH measurement system designed for water quality monitoring, hydroponics, aquariums, and educational projects. It consists of a plastic-shell combination pH electrode (E-201-C-9) and a signal conditioning module that converts the electrode’s high-impedance millivolt output into an analog voltage readable by Arduino, Raspberry Pi (with ADC), or other microcontrollers.
The E-201-C-9 electrode is a refillable combination electrode with a polycarbonate (PC) body and a glass bulb sensing membrane. It contains a Ag/AgCl reference electrode and uses a fiber liquid junction . The refillable design (unlike sealed gel electrodes) allows the user to replenish the 3 mol/L KCl reference solution when needed, extending electrode life . The electrode measures 0–14 pH over a 5–60°C range and terminates in a BNC connector .
The signal conditioning module provides a BNC socket for direct electrode connection and outputs an analog voltage proportional to pH. It operates from 5V DC with <10mA current and is compatible with 3.3V and 5V microcontroller ADCs . The module typically includes onboard potentiometers for zero (pH 7.00) and slope (pH 4.00/9.18) calibration, and some variants provide a DS18B20 temperature sensor interface for automatic temperature compensation .
Critical limitations: The electrode must be calibrated with standard buffer solutions before use—without calibration, readings are meaningless . The glass bulb is fragile and must not contact hard objects . The electrode should be stored wet in 3 mol/L KCl solution; drying out degrades performance . The module outputs an analog voltage, not a direct pH value—the microcontroller must perform the conversion. The electrode is not suitable for strong acids, strong bases, or fluoride solutions .
Features
-
Refillable Combination Electrode: Polycarbonate body with Ag/AgCl reference and fiber junction; can be replenished with 3M KCl .
-
0–14 pH Measurement Range: Suitable for most aqueous solutions .
-
BNC Interface: Standard connector for direct module attachment .
-
5V Analog Output Module: Outputs voltage proportional to pH; compatible with 3.3V/5V MCUs .
-
Onboard Calibration Potentiometers: Adjustable zero and slope for two-point calibration .
-
Temperature Compensation Ready: Some variants include DS18B20 interface for automatic compensation .
-
Compact PCB: Approx. 42×32mm module for easy integration .
Technical Specifications
| Parameter | Specification (E-201-C-9 Electrode) |
|---|---|
| Measurement Range | 0–14 pH |
| Temperature Range | 5–60°C (most sources); 0–80°C per one kit listing |
| Body Material | Polycarbonate (PC) |
| Reference Electrode | Ag/AgCl |
| Liquid Junction | Fiber |
| Internal Filling | Refillable (3M KCl) |
| Sensing Membrane | Glass bulb |
| Dimensions | φ12 × 120–165mm |
| Connector | BNC (Q9) |
| Parameter | Specification (Sensor Module) |
|---|---|
| Supply Voltage | 5V DC |
| Output Signal | Analog voltage (0–3V or 0–5V depending on variant) |
| Accuracy | ±0.1 to ±0.15 pH (at 25°C, after calibration) |
| Current | <10mA |
| Interface | BNC socket for electrode |
| Calibration | Onboard potentiometers for zero/slope |